10.
Secure the USB PCB by fitting the three flat washers (C), spring washers (D) and nuts
(E) retained in Step 4, in the locations shown in
Figure 12-7
.
Do not fit securing hardware to the top left-hand threaded spacer at this stage.
11.
Attach the USB cable (F) to the USB connector on the USB PCB.
12.
Refer to
Figure 7-16
. Fit the tag connector (C) to the top left-hand threaded spacer
(D). Secure the tag connector with the flat washer (F), spring washer (G), and nut (B)
retained in Step 6.
13.
Attach the spade connector at the free end of the Earth wire (A) to the tag connector
(C).
7.11
Replacing the power socket fuses
The analyzer contains two external fuses that are located in the fused power socket on the
rear panel.
Observe all safety precautions and allow the equipment to cool down before performing
this procedure.
The same procedure is used to replace both external fuses; therefore, the replacement of
only one is described.
DANGER!
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the power
ON/OFF
switch at the rear of the panel is set to
OFF
and tagged off before removing the top
cover.
The analyzer must be earthed.
Death or personal injury may result if this is not observed.
WARNING!
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and transporting
to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 31 kg (68 lb). Emerson
™
recommends that the analyzer is only moved and
lifted by a minimum of two people. Wear suitable protective gloves and protective footwear.
Failure to properly handle the analyzer may cause injury to personnel.
Troubleshooting and diagnostics
80
Rosemount CT5400
Summary of Contents for Rosemount CT5400
Page 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Page 16: ......
Page 20: ...Contents iv Rosemount CT5400 ...
Page 32: ...Plan 12 Rosemount CT5400 ...
Page 40: ...Install 20 Rosemount CT5400 ...
Page 44: ...Startup procedure 24 Rosemount CT5400 ...
Page 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Page 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Page 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Page 166: ...Preventative maintenance 146 Rosemount CT5400 ...