5.
Refer to
Figure 6-2
. Using
and
DOWN
, select GAS SERVICE.
6.
Press
.
The Select gas screen (
Figure 6-10
) opens.
Select gas Screen
Figure 6-10:
7.
Use
and
to move the cursor until the gas that you wish to span calibrate is
highlighted by the cursor. Press
.
The
Select type
screen (
Figure 6-11
) opens.
Select Type Screen
Figure 6-11:
8.
Use
and
to move the cursor until
SPAN
is selected (
Figure 6-11
). To perform a
calibration, press
.
The Mode selection screen (
Figure 6-12
) opens.
Gas Calibration Procedures
52
Rosemount CT5400