5.
Disconnect the Ethernet cable that links the Ethernet switch to the external Ethernet
port from Ethernet connector number 4 (C).
6.
Disconnect the Ethernet cable that links the Ethernet switch to the motherboard
from Ethernet connector number 5 (D).
7.
Disconnect the wiring harness (E) from the Ethernet switch.
8.
Press the small lever (F) on the underside of the Ethernet switch to release it from
the DIN rail. While continuing to press the lever, remove the Ethernet switch from
the DIN rail.
9.
Inspect the replacement Ethernet connector for damage. If damaged, contact your
local Customer Care Representative.
10.
Discard the unserviceable Ethernet switch.
11.
Place the replacement Ethernet switch (B) in its correct location on the DIN rail.
While pressing the small lever (E) on the underside of the Ethernet switch to the
down position, push the Ethernet switch onto the DIN rail. Release the small lever (E)
and check that the Ethernet switch is secure.
12.
Connect the wiring harness (F) to the Ethernet switch.
13.
On the Ethernet switch, connect the Ethernet cable from the motherboard to
Ethernet connector number 5 (D).
14.
Connect the Ethernet cable from the external Ethernet Port to Ethernet connector
number 4 (C).
15.
Connect the Ethernet cable from the analog input unit to Ethernet connector
number 1 (A).
16.
Refit the top cover as described in
Section 7.25
.
17.
Close the enclosure cover and secure with the twenty M16 bolts.
18.
Reinstall the analyzer in the rack. This step requires two people to properly lift and
transport it to the rack.
19.
Apply power to restart the analyzer.
7.17.5
Replacing the laser modules
The analyzer contains multiple laser modules; the precise number of laser modules is
determined by the configuration of the analyzer.
DANGER!
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the power
ON/OFF
switch at the rear of the panel is set to
OFF
and tagged off before removing the top
cover.
The analyzer must be earthed.
Death or personal injury may result if this is not observed.
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual
107
Summary of Contents for Rosemount CT5400
Page 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Page 16: ......
Page 20: ...Contents iv Rosemount CT5400 ...
Page 32: ...Plan 12 Rosemount CT5400 ...
Page 40: ...Install 20 Rosemount CT5400 ...
Page 44: ...Startup procedure 24 Rosemount CT5400 ...
Page 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Page 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Page 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Page 166: ...Preventative maintenance 146 Rosemount CT5400 ...