6.
Press
and
to list the fault listed.
7.
Press
to display the
Fault
diagnostic information screen.
Cell heater Temperature Out of Range
Figure 7-8:
8.
Examine the data shown on the Fault screen to determine if the suspect component
has failed or is operating outside of its correct parameters. If you do not have the
necessary information to determine if the suspect component is faulty, contact your
local Customer Care representative.
9.
From the Main menu, press
and
to scroll to
DIAGNOSTICS
.
Main Menu
Figure 7-9:
The Diagnostics screen lists those main components of the analyzer where problems
can be diagnosed using the BIT system. The Diagnostics screen also enables you to
check on the status and, where appropriate, the values of any of the listed
components.
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual
63