Analog IN/OUT Unit
Figure 7-25:
A. Analog I/O unit
B. Wiring harness connector (top)
C. Wiring harness connector (bottom)
D. Wiring harness connector (bottom)
5.
Disconnect the two wiring harness connectors (C and D) from the bottom of the
analog input unit.
6.
Release the analog input unit from the DIN rail by pressing a small lever on the
underside of the unit. While continuing to press the lever, remove the analog input
unit from the DIN rail.
7.
Discard the unserviceable analog input unit.
8.
Inspect the replacement unit for damage. If damage has occurred, contact your
local Emerson Customer Care Representative.
9.
Insert the replacement analog input unit in its correct location on the DIN rail. Press
the small lever on the underside of the unit and push the unit onto the DIN rail.
Release the small lever and check that the unit is secure.
10.
Connect the two wiring harness connectors (C and D) to the bottom of the unit.
11.
Refit the top cover as described in
Section 7.25
.
12.
Remove the protective masking tape from the flanges, seal,and flamepath.
13.
Remove Lock-Out Tag-Out labels.
14.
Apply power to the analyzer.
Troubleshooting and diagnostics
100
Rosemount CT5400
Summary of Contents for Rosemount CT5400
Page 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Page 16: ......
Page 20: ...Contents iv Rosemount CT5400 ...
Page 32: ...Plan 12 Rosemount CT5400 ...
Page 40: ...Install 20 Rosemount CT5400 ...
Page 44: ...Startup procedure 24 Rosemount CT5400 ...
Page 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Page 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Page 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Page 166: ...Preventative maintenance 146 Rosemount CT5400 ...