The equipment shall only be used with process gases which are in gas groups B, C or
D (Divisions) or IIB + H
2
(Zones) and must not contain oxygen or any other oxidizer in
concentrations greater than that found in normal air.
•
This assessment does not cover reliable function, performance, or other properties
of the equipment not related to safety.
•
The equipment is to be installed using wire no larger than 1mm
2
(18 AWG).
•
The equipment is only to be installed by manufacturer trained personnel.
•
If at any time there is a conflict between the system safety provisions and any
relevant local (national or regional) requirements, local requirements always take
precedence.
•
The equipment is not to be used with flammable liquids.
1.6
Qualified personnel
This manual provides installation, operation, and maintenance personnel with the level of
knowledge required to safely start, operate, and switch off the analyzer.
Only technically qualified personnel in the field of instrumentation and control who are
familiar with this manual and have been specially trained on the analyzer should install,
operate, switch off, and service the analyzer. Only qualified and trained persons have the
required specific knowledge to correctly interpret the general safety information,
warnings, and procedures given in this manual and apply them to this particular
application. Emerson or its distribution partners can provide this training on request.
Knowledge of the safety information within this manual and its technically correct
implementation are prerequisites for danger-free installation, operation, and maintenance
of the analyzer.
1.7
Software version
The analyzer includes software that is used to control the operation of the instrument. This
manual describes software version 5.x.x.
1.8
Gas detection
The analyzer is highly configurable in the gases that can be detected and their range of
concentrations.
1.9
Detailed system specifications
Table 1-2
gives the physical characteristics of the analyzer. Schematic diagrams of the
sensor and mounting points are shown in
Figure 1-3
,
Figure 1-4
, and
Figure 1-5
.
Table 1-3
gives the general characteristics of the instrument.
Plan
6
Rosemount CT5400
Summary of Contents for Rosemount CT5400
Page 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Page 16: ......
Page 20: ...Contents iv Rosemount CT5400 ...
Page 32: ...Plan 12 Rosemount CT5400 ...
Page 40: ...Install 20 Rosemount CT5400 ...
Page 44: ...Startup procedure 24 Rosemount CT5400 ...
Page 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Page 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Page 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Page 166: ...Preventative maintenance 146 Rosemount CT5400 ...