6
Gas Calibration Procedures
6.1
Required tools
To calibrate the analyzer, you need the following items:
•
Nitrogen gas of
instrument gas
purity for use as a zero calibration gas
•
Suitable span calibration gases for each gas measured
•
Gas bottle pressure regulators
•
Interconnecting hoses to connect the gas bottles to the analyzer
•
A T-piece and excess flow line
NOTICE
In the case of gases such as H
2
O, where it is not normally possible to obtain calibrated gas
cylinders, it will usually be measured by the same laser as some other gas. Validating the other
gases measured by the analyzer (particularly any which are measured by the same laser as
H
2
O) can demonstrate that the system is functioning correctly, meaning there is no need to
calibrate the water measurement directly.
If calibration must be carried out (e.g., for legal requirements), it can be calibrated by using a
water vapor generator to supply a known concentration of water vapor.
6.2
Main menu
The calibration functions are accessed through the
Main menu
(
Figure 6-1
). To get to the
Main menu
, press
MENU
on either the
Gas Sensor Main
screen or the
Pressure and
Temperature
screen, as described in
Section 4.9
.
Main menu
Figure 6-1:
Gas Calibration Procedures
Installation, Operations, and Maintenance Manual
45
Summary of Contents for Rosemount CT5400
Page 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Page 16: ......
Page 20: ...Contents iv Rosemount CT5400 ...
Page 32: ...Plan 12 Rosemount CT5400 ...
Page 40: ...Install 20 Rosemount CT5400 ...
Page 44: ...Startup procedure 24 Rosemount CT5400 ...
Page 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Page 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Page 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Page 166: ...Preventative maintenance 146 Rosemount CT5400 ...