Mode selection screen
Figure 6-12:
9.
Press
to perform a manual calibration.
The Span input screen (
Figure 6-13
) opens.
Span input Screen
Figure 6-13:
10.
Use
and
to highlight each digit in turn; then use
and
to increase or
decrease the value until the concentration displayed matches the cylinder you are
using. Press
to proceed to the next step.
NOTICE
The cylinder concentration must be entered in ppm.
The begin Calibrate span screen (
Figure 6-14
) opens.
Gas Calibration Procedures
Installation, Operations, and Maintenance Manual
53