17.
If the calibration of the analyzer remains outside of tolerance, you can adjust the
calibration by pressing
.
This brings up a Calibrate screen (
Figure 6-16
).
Calibrate - re-calibrate with New ZERO offset screen
Figure 6-16:
18.
To proceed with the automatic re-calibration, press
.
The automatic re-calibration of the instrument's span calibration starts, and the
offset shown on the screen (in
Figure 6-17
it is 0.997) is applied to the instrument.
Press
or
to abort the calibration process.
When the calibration or verification is complete, the screen shown in
Figure 6-17
is
displayed.
Calibrate Complete Screen
Figure 6-17:
19.
If you wish to perform a span calibration for another gas, press
.
The software returns to the Select gas screen (
Figure 6-10
). Repeat steps 7-18 for the
next gas.
Gas Calibration Procedures
Installation, Operations, and Maintenance Manual
55