Calibration / Verification Screen
Figure 5-14:
21.
If you wish to perform a span for another gas, press
.
The software returns to the
Select gas
screen (
Figure 5-1
) Repeat steps 8 - 16 for the
next gas.
22.
If you wish to perform a zero verification, press
. Then follow the zero verification
procedure in
Section 5.1.1
.
23.
If you are finished verifying the analyzer, press
.
The software returns to the
Gas Sensor Main
screen (
Figure 4-7
).
The span verification procedure is now complete.
Verifying Gas Concentrations
44
Rosemount CT5400