Power Socket Fuses
Figure 7-17:
A. Fuse holder
B. Fuse
C. Cover
4.
Pull out the fuse holder (A).
5.
Remove the fuse (B) from the holder (A) .
6.
Fit a new fuse (B), type: 240 V, 50 Hz, 500 mA, fast-acting, into the fuse holder (A).
7.
Fully push in the fuse holder (A).
8.
Close the fuse cover (C).
9.
Reconnect the external power supply to the power socket.
10.
Start up the analyzer by turning the
ON/OFF
switches to the
ON
position.
7.12
Replacing the internal fuses
The analyzer contains a number of internal fuses. The same procedure is used to replace all
of the internal fuses; therefore, the replacement of only one is described.
Troubleshooting and diagnostics
82
Rosemount CT5400