Behind the second coupling echo, artefacts have been produced during the coupling procedure
(probably multiple reflections in the couplant layer). For that reason, it was switched to domain
clipping. The first clipping domain has been defined in a way to suppress the afore mentioned
artefacts (see fig. on the right). The second clipping domain is operated in the global clipping mode
with the amplitude limit set to 0 and ranging until the end of measuring range.
Five layer interface echoes have been extracted, hence five layer thickness values have been
provided as a result.