© ElektroPhysik
QuintSonic 7
101 von 184
The current batch will be associated with the parameter set you have
chosen. The name of the new parameter set is shown in the
“Configuration” submenu.
Option 2: via the measure mode screen
From the measure mode screen press function key “Configuration” to
call this submenu.
From this point proceed according to option 1.
Note:
This function mainly serves to assign a parameter to a new batch other than the “Default
parameter set”. Make sure the new batch does NOT contain any readings.
The parameter set of a batch containing readings should never be changed. Any changes to its
original parameter set would change the conditions for measurement.
As a result, the batch would contain a mixture of readings taken with the originally set parameters
as well as readings taken with the new parameter set. Hence, consistency of this batch would not
be given any more.