© ElektroPhysik
QuintSonic 7
114 von 184
12.1.4.4
Ultrasound velocity
The system supports up to 5 layers. The sound velocity of layers can
be adjusted in this menu. To enter sound velocity, several options are
available (see 5.2 “How to calibrate for the sound velocity”).
Use arrow up/down keys to select the requested layer. It is of no
significance whether “Material” or “Sound velocity” are marked.
Press OK to confirm and call the entry method.
Entry options:
Direct (numerical entry)
Literature values (selection from a sound velocity table)
User values
Determine (Calibration): via target layer thickness and thickness
measurement
Make your selection and press OK to confirm and call the respective
submenu.
12.1.4.4.1 Direct entry
Select “Direct” to call a numerical entry field. Enter the value for sound
velocity. Unit of measure is m/s. For entry of alphanumerical para-
meter values please refer to section 4.3.4.