© ElektroPhysik
QuintSonic 7
154 von 184
15.6.4.2
Data output setting “RS232 cable”
Chose “RS232“ if your computer has no IR interface or if the QuintSonic
7 is to be permanently connected to a PC.
The baud rate can be adjusted as requested form 57600 (default),
38400, 19200 to 9600.
The data format
“8D/1S/N“ (8 data bits, 1 Stop-Bit, No Parity) is preset
and cannot be changed.
Under “Transmission protocol” you can enable or disable the
ElektroPhysik transmission protocol.
Set to “ON” if you are using QSoft 7.
Set to “OFF” if data transfer is made to another software. In this setting,
data transfer is made in ASCII format with carriage return character
(CR/LF). See table below.
Description
Ref. # of reading
Space
Reading
Measuring
system
Carriage
Return
Line
Feed
Note
Only if storage of
readings into the
statistics memory is
enabled
aligned to the right
only if fixed point
format is disabled
Example
Reading
147 104.8µm
0x31 0x34 0x37
0x20
0x31 0x30 0x34 0x2E
0x38
0xB5 0x6D
0x0D
0x0A