Chapter 5 Examining Spectrum Data
5-54
Applied Biosystems
5
General
guidelines for
setting
parameters
Refer to the following table to determine how to set Advance
Baseline Correction parameters and obtain the desired
baseline correction.
Condition
Set Parameters To
Baseline follows a rising peak cluster
•
Peak Width—Average peak width
•
Flexibility—0.5
•
Degree—0.5
Baseline is gently curving
•
Peak Width—10 to 20 times peak width
•
Flexibility—0
•
Degree—0.5 (decrease further to raise
the baseline)
Baseline rise should be ignored and
treated as signal
•
Peak Width—Half of the width of the
cluster to correct (in the illustration
shown, if the width of the cluster is 200,
set Peak Width to 100)
•
Flexibility—0
•
Degree—0.5 (decrease further to raise
the baseline)
Peak cluster
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...