Mass Deconvolution (Mariner Data Only)
Data Explorer
™
Software User’s Guide
5-37
5
5.6 Mass Deconvolution
(Mariner Data Only)
NOTE: Mass deconvolution is not supported for
Mariner DAD data.
NOTE: The Mass Deconvolution software is an option in the
Data Explorer software. The Multiple Charge command on
the Process menu is dimmed if you have not purchased the
option.
Overview
You can use the Data Explorer Mass Deconvolution features
to generate a theoretical zero-charge spectrum that
represents the molecular mass of a protein:
•
Mass Deconvolution—Use when you have a
spectrum with clearly resolved, multiply charged peaks.
You specify the m/z values for the peaks to include.
This function requires at least two adjacent peaks within
the same charge envelope.
•
Convert to Zero-Charge Spectrum—Use when you
have a spectrum with overlapping charge envelopes, or
a noisy baseline. You specify a mass range for the
peaks to include and a mass range for the molecular
mass.
Mass
Deconvolution
To use the Mass Deconvolution function:
1.
Click the Spectrum window.
2.
Select Duplicate Active Trace from the Display menu to
keep the original data displayed after processing.
3.
From the Process menu, select Multiple Charge, then
select Mass Deconvolution.
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...