Chapter 5 Examining Spectrum Data
5-52
Applied Biosystems
5
Flexibility
With the Peak Width parameter, determines the number
of points used to estimate the baseline amplitude at
regularly spaced points in the spectrum, but is not directly
proportional to the number of points used. Valid entries
are 0 to 1.0.
The default value of 0.5 works best for most applications.
A value closer to 0 reduces flexibility and provides a
smoother more generalized baseline correction (see
below).
A value closer to 1 provides more localized baseline
correction by using a larger number of points to estimate
the baseline (see below).
If the baseline rises adjacent to peaks after
baseline-correcting, decrease the Flexibility value and
baseline-correct again.
(continued)
Parameter
Description/Specifies
Flexibility=1
Flexibility=0
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...