Chapter 3 Peak Detection and Labeling
3-26
Applied Biosystems
3
Peak Processing
parameters
(spectrum data
only)
Table 3-3 describes the parameters in the Peak Processing
tab of the Spectrum Peak Detection Setup dialog box (see
Figure 3-5 on page 3-16).
Default peak detection values are listed in Section 3.7, Default
Peak Detection Settings.
Table 3-3 Peak Processing Parameters (Spectrum Data Only)
Parameter
Description
Integration Baseline Settings
NOTE: You can set peak labels to display peak start, peak end, and baseline
marks. See Section 3.5.2, Setting Chromatogram and Spectrum Peak Labels.
Valley-to-Baseline
Draws a vertical line from all valleys to a horizontal baseline.
The level of the horizontal baseline is determined using the
minimum peak valley point (left or right) for each peak. See
Figure 3-25 on page 3-70.
Valley-to-Valley
Forces a baseline through all valley points. See Figure 3-25
on page 3-70.
Spectrum Parameters
Centroid %
Specifies the percentage of peak height used to determine
the location of the centroid mass. For example, with a
Centroid % of 10, the software examines the top ten percent
of signal.
Enable Gaussian
Fitting
Increases the accuracy of the centroid calculation for
spectral peaks with limited data points (fewer than 6 data
points across the peak), in particular, Mariner data below
m/z 400.
Continued
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...