Overview
Data Explorer
™
Software User’s Guide
3-3
3
3.1.2 The Resolution-Based
Peak Detection Routine
This section describes:
•
Type of data affected
•
Process that occurs
•
Detection ranges
•
PSD peak detection for Voyager data
Type of data
affected
The resolution-based peak detection routine applies to
spectral data only. You can enable or disable the
resolution-based peak detection routine as described on
page 3-14.
Process that
occurs
When enabled, the resolution-based peak detection routine
automatically:
•
Divides a trace into detection ranges based on the
expected number of data points across a typical mass
spectral peak.
•
Applies a Filter Width that is equal to the number of data
points per peak in each detection range.
•
Uses a Filter Increment of 1.
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...