Peak Detection
Data Explorer
™
Software User’s Guide
3-17
3
Setting Advanced
Settings
(spectrum data
only)
To set Advanced Settings that you can apply locally to a
selected detection range, and that override the thresholds set
in the Basic Settings tab:
1.
In the Basic Settings tab, select Use Advanced
Settings.
2.
Click the Advanced Settings tab in the Spectrum Peak
Detection Setup dialog box.
NOTE: If you select Use Resolution Dependent
Settings in the Basic Settings tab, Basic Settings
override Advanced Settings. The Advanced Settings
tab is accessible, but all parameters are dimmed. To
make Advanced Settings available for editing, select
Use Advanced Settings on the Basic Settings tab.
The Advanced Settings tab is displayed (Figure 3-6).
Figure 3-6 Spectrum Peak Detection Setup—
Advanced Settings Tab
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...