Chapter 6
Performing the Pure Dye Calibration
Creating a Plate Document for the Pure Dye Calibration
72
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Creating a Plate Document for the Pure Dye Calibration
1.
In the SDS software, click
(or
select
File New
).
2.
Configure the New Document dialog box:
a.
Select
Assay Pure Spectra
.
b.
Select
Container 96-Well Clear
.
c.
Select
Template Blank Document
.
d.
In the Operator field, enter your name.
e.
In the Comments field, enter any
information that you want to attach to the
file (such as the plate bar code).
f.
Click
.
Note:
It is not necessary to name or save
the pure dye plate document. The SDS
software automatically saves the pure dye
data to a calibration file on the computer
hard drive.
The Pure Spectra Calibration Manager is
displayed.
“Preparing Plates and Performing the
Pure Dye Calibration” on page 73
.
a
b
c
d
e
f