Chapter 1
Before Starting the Installation
Planning the Installation
6
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
~ 02:00:00
7300/7500
Perform the pure dye calibration
(Chapter 6):
a. Prepare the
pure dye
plates.
b. Create a plate document.
c. Perform the
pure dye
calibration.
d. Analyze the pure dye calibration data.
e. Evaluate pure dye calibration spectra.
~ 01:20:00
~ 01:00:00
Verify instrument performance
(Chapter 7):
a. Prepare the TaqMan
®
RNase P verification plate.
b. Create a plate document.
c. Perform the
verification run
.
d. Analyze the TaqMan
®
RNase P plate data.
7500 Fast
~ 00:40:00
7500/7500 Fast
7300
Note: You must be present at the instrument for approximately 5 minutes of the total run time.
7500 Fast
Recommended Workflow (continued)
See Note below
Perform the optical calibration
(7500/7500 Fast systems only,
Chapter 5):
a. Create a plate document.
b. Perform the optical calibration.
c. Analyze the optical calibration data.
~ 00:10:00
Perform the background calibration
(Chapter 5):
a. Prepare the
background
calibration plate.
b. Create a plate document.
c. Perform the background calibration.
d. Analyze the background data.
~ 00:30:00
Time Required
Task