Appendix A
Decontaminating the Sample Block
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
111
Notes
A
9.
Pull the heated cover door to the front of the
instrument, then close the access door.
10.
Plug in, then power on the 7300/7500/7500 Fast
system.
11.
Create a new background plate (see
.)
12.
To confirm that you have eliminated the
contamination, perform a background calibration
run (see
Calibration and Optical Calibration” on
page 49
13.
If the contamination is still present, repeat
.
CHEMICAL HAZARD.
Ethanol
is a flammable liquid and vapor.
Exposure causes eye, skin, and respiratory tract
irritation and may cause central nervous system
depression and liver damage. Read the MSDS,
and follow the handling instructions. Wear
appropriate protective eyewear, clothing, and
gloves.
Contamination