Chapter 4
Performing the Regions of Interest (ROI) Calibration
Creating a Plate Document for the ROI Calibration
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
41
Notes
4
Creating a Plate Document for the ROI Calibration
1.
Click
(or
select
File New
).
2.
In the New Document wizard, click
to
accept the default parameters.
Note:
It is not necessary to name or save the
plate document. The SDS software automatically
saves the ROI data to a set of calibration files on
the computer hard drive.
If a Network communications error is displayed,
click
. You can perform the ROI
calibration even if this error is displayed.
Note:
This message is displayed if your
computer is not connected to a network. If you
do not plan to connect this computer to a
network, select
Do not show me this message
again
, then select
Never again on this machine
.
If an ROI error is displayed, click
to
proceed with ROI calibration.
Select
Click
Click
Click