Chapter 3
Installing the Software and Powering On the Instrument
Troubleshooting
34
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Condition: CCD, Filter Wheel, or Shutter Test Failures
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see
).
3. If the test fails again, perform the ROI calibration as explained
in
“Performing the Regions of Interest (ROI) Calibration” on
.
4. Determine if the ROI image appears in the ROI Inspector.
Yes
– Continue with the installation.
No
– Contact Applied Biosystems technical support (see
) or your service representative.
Condition: Lamp Test Failure
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see
).
3. If the test fails again, select
Instrument
Lamp
Status/Replacement
.
4. If the Lamp Status/Replacement dialog box reports the lamp
status as Failed, replace the halogen lamp (see
5. Perform the Instrument Function Test (see
).
6. If the test fails again, contact Applied Biosystems technical
support (see
) or your service representative.
Condition: Thermal Cycler Test Failure
1. Power off the instrument, wait 30 seconds, then power on the
instrument.
2. Perform the Instrument Function Test (see
).
3. If the test fails again, contact Applied Biosystems technical
support (see
) or your service representative.
Troubleshooting – Function Tests
(continued)
Lamp
status