Chapter 8
Powering Off the 7300/7500/7500 Fast System
Performing a Long-Term Shutdown
104
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Performing a Long-Term Shutdown
Perform the long-term shutdown procedure if the
instrument will be inactive for more than 7 days.
Time Required
5 minutes
Materials Required
Performing a Long-term Shutdown
1.
Press the tray to open it.
2.
If the tray contains a plate, remove it.
3.
Load the packaging plate into the tray.
Note:
If the shipping plate is not available,
substitute an unused reaction plate. During
storage, the instrument optics block rests on the
plate to protect the optics block.
4.
Press the tray to move it into the instrument.
5.
Press the instrument power button.
6.
Power off the computer and monitor:
a.
Select
Shut Down
.
b.
In the Shut Down Windows dialog box (not
shown), select
Shut Down
.
c.
Power off the monitor.
Su M T W Th F S
> 1 week
Load
Summary of Contents for 7300