Chapter 4
Performing the Regions of Interest (ROI) Calibration
Overview
36
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Overview
Note:
You can also use the Real-Time System Setup wizard to perform the procedure in
this chapter. To access the wizard, select
Program Files 7300/7500/7500
Fast System
Real-Time System Setup Wizard
.
Time Required
30 minutes
Materials
Required
When to Perform
the ROI
Calibration
Perform a regions of interest (ROI) calibration:
• When installing the system (you must perform all calibrations in this order: ROI
background optical (7500/7500 Fast systems only) pure dye instrument
verification).
• Every 6 months, or as often as necessary, depending on instrument use.
• After replacing the lamp.
IMPORTANT!
After every ROI calibration, you must perform a background calibration,
optical calibration (7500/7500 Fast systems only), pure dye calibration, and instrument
verification.
GR2170
Powder-free
Gloves
Safety
Goggles
ROI Calibration Plate
Centrifuge with plate
adapter
Applied Biosystems
Real-Time PCR System
Spectral Calibration Kit