5.2. Specifications
Infant
Pediatric
Adult
Resuscitator volume
approx. 220 ml
approx. 670 ml
approx. 1510 ml
Delivered volume one hand*
150 ml
450 ml
600 ml
Delivered volume two hands*
-
-
1000 ml
Dimensions
(length x diameter) w/o
reservoir and accessory
approx.
190 x 71 mm
approx.
223 x 99 mm
approx.
284 x 127 mm
Weight w/o reservoir
and accessory
approx. 70 g
approx. 145 g
approx. 220 g
Pressure-limiting valve**
4.0 kPa
(40 cm H
2
O)
4.0 kPa
(40 cm H
2
O)
4.0 kPa
(40 cm H
2
O)
Dead space
≤ 5 ml + 10 % of the
delivered volume
≤ 5 ml + 10 % of the
delivered volume
≤ 5 ml + 10 % of the
delivered volume
Inspiratory resistance***
max 0.1 kPa
(1.0 cm H
2
O)
at 5 l/min
max 0.5 kPa
(5.0 cm H
2
O)
at 50 l/min
max 0.5 kPa
(5.0 cm H
2
O)
at 50 l/min
Expiratory resistance***
max 0.2 kPa
(2.0 cm H
2
O)
at 5 l/min
max 0.27 kPa
(2.7 cm H
2
O)
at 50 l/min
max 0.27 kPa
(2.7 cm H
2
O)
at 50 l/min
Reservoir volume
approx. 300 ml (bag)
approx. 100 ml (tube)
approx. 2600 ml (bag) approx. 2600 ml (bag)
Patient connector
Outside 22 mm male (ISO 5356-1)
Inside 15 mm female (ISO 5356-1)
Expiration connector
(for PEEP valve attachment)
30 mm male (ISO 5356-1)
Manometer Port connector
Ø 4.2 +/- 0.1 mm
Demand Valve connector
-
Inside 32 mm female (EN ISO 10651-4)
Infant
Pediatric
Adult
Forward and
backward leakage
Not measurable
M-Port
Connector compatible with EN ISO 80369-7
O
2
inlet connector
According to EN 13544-2
Operation temperature limits -18 °C to +50 °C (-0.4 °F to +122 °F), tested according to EN ISO 10651-4
Storage temperature limits
-40 °C to +60 °C (-40 °F to +140 °F), tested according to EN ISO 10651-4
Recommended long term storage in closed packaging at room temperature, away from sunlight.
* Tested according to EN ISO 10651-4.
** Higher delivery pressure can be obtained by overriding the pressure-limiting valve.
*** At general test conditions according to EN ISO 10651-4.
5.3. MRI Safety Information
The Ambu SPUR II Resuscitator
1
, and Ambu SPUR II Resuscitator with attached Ambu
Disposable PEEP 20 Valve
2
, and Ambu SPUR II Resuscitator with attached Ambu Disposable
PEEP 20 Valve and Ambu Disposable Pressure Manometer
2
are MR Conditional, and
therefore may be safely used in the MR environment (not inside the MR bore) under the
following conditions.
Static magnetic field of 7 Tesla and less, with
• Maximum spatial field gradient of
– 10,000 G/cm (100 T/m)
1
– 16,000 G/cm (160 T/m)
2
• Maximum force product of
– 450,000,000 G2/cm (450 T
2
/m)
1
– 721,000,000 G2/cm (721 T
2
/m)
2
Use inside the MR bore may influence MR image quality.
RF-induced heating and MR image artifacts have not been tested. Any metallic parts are
fully encapsulated and do not have any contact with the human body.
Summary of Contents for SPUR II
Page 1: ...INSTRUCTIONS FOR USE Ambu SPUR II Disposable...
Page 16: ...16 4 2 2 2 2 1 2 2 0 C 0 C 10 l min 10 l min M Port M Port M Port 2 3 4 3 3 1 3 2...
Page 48: ...48 1 5 PEEP 4 2 4 2 1 4 2 2 0 C 4 2 2 2 2 1 2 2 0 C 0 C 10 l min 10 l min 2 3 4 3 3 1...
Page 170: ...170 1 4 1 2 3 Ambu PEEP 4 5 6 7 8 9 10 11 CO2 12 O2 13 EtCO2 400 EtCO2 14 1 15 O2 16 17 18 19...