Starting a Measurement (Trigger) and Detecting the Completion of a Measurement
(End of Sweeps)
Triggering a Measurement (Starting a Measurement)
1. Ch
apter
T
itle
2
. Cha
p
te
r T
itle
3
. Chap
te
r T
itle
4
. Chap
te
r T
itle
5. S
tart
ing a
Measur
ement(
T
rigger
)
and D
e
tecting t
he Completio
n
of a
Measure
m
ent(End of
S
w
eep
s)
The state transition after the completion of each sweep varies depending on the sweep
count setting.
Once
Transitions to the “Idle” state (
a
in Figure 5-1).
Specified number
of times
Repeats a transition to the “Waiting for Trigger” state until the
specified number of sweeps are completed (
b
in Figure 5-1).
After the completion of the specified number of sweeps, there is a
transition to the “Idle” state (
a
in Figure 5-1).
Continuous
Repeats a transition to the trigger wait state (
b
in Figure 5-1).
Triggering a measurement
To make measurements continuously (to automatically generate triggers
continuously)
Step 1.
Use the “TRGS” command to set the trigger source to the internal trigger.
Step 2.
Use the “CONT” command to set the sweep setting to continuous sweep.
To make a measurement at any desired time (to generate a trigger from an external
controller)
Step 1.
Use the “TRGS” command to set the trigger source to the internal trigger.
Step 2.
Use the “SING” or “NUMG” command to execute a single sweep or the specified number
of sweeps at any desired time.
Or
Step 1.
Use the “TRGS” command to set the trigger source to the GPIB/LAN trigger.
Step 2.
Use the “CONT” command to set the sweep setting to continuous sweep.
Step 3.
Execute the “*TRG” command on page 261 or the HP BASIC TRIGGER command
(execute the group execution trigger) at any desired time.
NOTE
You can use the group execution trigger (GET) when you want to simultaneously trigger
all devices connected to the controller that can be triggered.
Trigger source setting
Method of generating a trigger
Internal trigger (INT)
Automatically generates an internal trigger.
External trigger (EXT)
Enters a trigger signal from the EXT TRIGGER
terminal on the rear panel to generate a trigger.
GPIB/LAN trigger (BUS)
Executes the “*TRG” command on page 261 or the
HP BASIC TRIGGER command to generate a trigger.
Manual trigger (MAN)
Uses the following key sequence on the front panel to
generate a trigger:
[Trigger]
-
SOURCE[ ]
-
MANUAL
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...