Chapter 16
351
GPIB Command Reference
MKRAMPO
16. GPI
B
Com
m
an
d Re
fere
nce
( F -
N )
MKRAMPO
Syntax
MKRAMPO
Description
For the active trace (set with the “TRAC” command on page 457), sets the measurement
parameter value at the marker position as the offset value of the limit for the measurement
parameter value. (No query)
Corresponding
key
[System]
-
LIMIT TEST
-
LIMIT LINE OFFSETS
-
MKR
→
AMP. OFST.
MKRAUV?
Syntax
MKRAUV?
Description
For the active trace (set with the “TRAC” command on page 457), if the measurement
parameter is a vector value (for COMPLEX Z-Y), reads out the subsidiary readout of the
measurement parameter value at the marker position. (Query only)
Query response
{numeric}<newline><^END>
If the measurement parameter is a scalar value (for other than COMPLEX Z-Y), 0 is
always read out. If the measurement parameter is a vector value (for COMPLEX Z-Y), the
value is as follows.
•
If the polar chart format is selected, the phase is read out.
•
If the complex plane format is selected, reactance (for trace A) or susceptance (for trace
B) is reads out.
Corresponding
key
No front panel key is available to execute this function.
MKRCENT
Syntax
MKRCENT
Description
For the active trace (set with the “TRAC” command on page 457), sets the sweep
parameter value at the marker position as the center value of the sweep range. To set the
center value of each segment when creating the list sweep table, also use this command. If
the marker function is off, executing this command causes an error and the command is
ignored. (No query)
Corresponding
key
[Marker
→
]
-
MKR
→
CENTER
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...