462
Chapter 16
GPIB Command Reference
WIDVTYPE
WIDVTYPE
Syntax
WIDVTYPE {DIVS2|MULS2|DIV2|FIXed}
WIDVTYPE?
Description
For the active trace (set with the “TRAC” command on page 457), selects the method to set
a cutoff point in the trace bandwidth analysis function.
Parameters
If the measurement parameter is a vector value (for COMPLEX Z-Y), you can select
FIXed only.
Query response
{DIVS2|MULS2|DIV2|FIXed}<newline><^END>
Corresponding
key
[Search]
-
WIDTHS DEF MENU
- {
MKRVAL/(
√
2)
|
MKRVAL*(
√
2)
|
MKRVAL/2
|
FIXED VALUE
}
WOPEN
Syntax
ROPEN <string>
Description
If the specified file exists, this command makes it write-enabled; otherwise, creates a new
file and makes it write-enabled. This command takes its arguments in a different way,
depending on the file format.
The size of an existing file cannot be changed. Therefore, if you want to change them,
delete the file itself using the “PURG”(406page) command and then create a new file using
this command.
This command is used in combination with the “WRITE”(463page) command and the
“CLOSE”(280page) commands, as shown in Figure 16-8 on page 412. (No query)
Parameters
Corresponding
key
No front panel key is available to execute this function.
Description
DIVS2
The value obtained by dividing the measurement parameter value
at the marker position by
is set as the cutoff point.
MULS2
The value obtained by multiplying the measurement parameter
value at the marker position and
is set as the cutoff point.
DIV2
The value obtained by dividing the measurement parameter value
at the marker position by 2 is set as the cutoff point.
FIXed (initial
value)
Lets you set the cutoff point freely using the “WIDFVAL”
command on page 460.
2
2
Description
<string>
File name of up to 12 characters including its extention
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...