Chapter 16
301
GPIB Command Reference
DISP
16. GPI
B
Com
m
an
d Re
fe
re
nce
( * , A - E
)
DISP
Syntax
DISP {DATA|MEMO|DATM|DMNM|DELP|DDVM}
DISP?
Description
For the active trace (set with the “TRAC” command on page 457), makes the setting
related to the displayed trace.
If a parameter other than DATM is set, the scale setting target trace (set with the “SCAF”
command on page 422) and the marker use target trace (set with the “MKRO” command
on page 356) are also changed.
If no data has been copied to the memory trace (the “DATMEM” command on page 285
has not been executed), executing this command to set a parameter other than DATA
causes an error and the command is ignored.
If the measurement parameter is a scalar value, executing this command to set DDVM
causes an error and the command is ignored. If the measurement parameter is a vector
value, executing this command to set DELP causes an error and the command is ignored.
Parameters
Query response
{DATA|MEMO|DATM|DMNM|DELP|DDVM}<newline><^END>
Corresponding
key
[Display]
-
DEF TRACE[ ]
- {
DATA
|
MEMORY
|
DATA and MEMORY
|
DATA-MEM
|
DELTA %
|
DATA/MEM
}
Description
*1
DATA
(initial value)
Displays the data trace. The contents of the data trace is measured
data (no operation).
MEMO
*2
Displays the memory trace.
DATM
*2
Displays both the data trace and the memory trace at the same time.
The contents of the data trace is measured data (no operation).
DMNM
*2
Displays the data trace. The contents of the data trace is the
operation result of DATA - MEM.
DELP
*2,*3
Displays the data trace. The contents of the data trace is the
operation result of (DATA - MEM) / MEM.
DDVM
*2,*4
Displays the data trace. The contents of the data trace is the
operation result of DATA / MEM.
*1.In operation expressions in the description, DATA indicates measured
data and MEM indicates data of the memory trace.
*2.Selectable only when data has been copied in the memory trace.
*3.Selectable only when the measurement parameter is a scalar value.
*4.Selectable only when the measurement parameter is a vector value.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...