Chapter 16
381
GPIB Command Reference
OUTPIACP?
16. GPI
B
Com
m
an
d Re
fe
re
nce
( O -
R )
OUTPIACP?
Syntax
OUTPIACP? <numeric 1>
Description
When the oscillator level monitor function is set to on (set to ON with “OMON” command
on page 365), reads out the oscillator current level monitor value of the specified
measurement point. (Query only)
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric 2}<newline><^END>
If the oscillator level monitor function is set to off, the readout is invalid.
Corresponding
key
No front panel key is available to execute this function.
OUTPINPCIO?
Syntax
OUTPINPCIO?
Description
When input/output port C of the 24-bit I/O port is set to an input port (by the execution of
the “CIN” command on page 278), 4-bit data is inputted into the 4294A through port C (C0
to C3) and the data is read out to the external controller. If port C is set to an output port,
executing this command causes an error and the command is ignored. This command has
the same function as the READIO(16,2) command of Instrument BASIC.
Query response
{numeric}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric 1>
Description
Number of the measurement point you want to read out (the
number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
Range
1 to the number of points
Resolution
1
{numeric 2}
Description
Current monitor value of the oscillator level
{numeric}
Description
Decimal value for the 4-bit binary data inputted from C0-C3,
assuming that C0 is LSB and C3 is MSB.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...