Chapter 16
383
GPIB Command Reference
OUTPLIMF?
16. GPI
B
Com
m
an
d Re
fe
re
nce
( O -
R )
OUTPLIMF?
Syntax
OUTPLIMF?
Description
For the active trace (set with the “TRAC” command on page 457), reads out the limit test
results of measurement points that failed the test (FAIL). (Query only)
Query response
{numeric 1},{numeric 2},..,{numeric F_NOP
×
4}<newline><^END>
Where, F_NOP is the number of points that failed the test, which can be obtained using the
“OUTPFAIP?” command on page 380. As shown below, for each measurement point, NOP
sets of values (sweep parameter value, test result, upper limit, and lower limit) are read out
in this order (total of F_NOP
×
4 values).
{numeric n
×
4+1}: Sweep parameter value
{numeric n
×
4+2}: Limit test result (always 0)
{numeric n
×
4+3}: Upper border value of the limit
{numeric n
×
4+4}: Lower border value of the limit
Where, n is an integer between 0 and F_NOP - 1.
Note that if there is no failed measurement point, only 1 piece of data, whose value is 1, is
read out.
Corresponding
key
No front panel key is available to execute this function.
OUTPLIML?
Syntax
OUTPLIML?
Description
For the active trace (set with the “TRAC” command on page 457), reads out the limit test
results of all measurement points. (Query only)
Query response
{numeric 1},{numeric 2},..,{numeric NOP
×
4}<newline><^END>
Where, NOP is the number of points specified using the “POIN” command on page 401.
As shown below, for each measurement point, NOP sets of values (sweep parameter value,
test result, upper limit, and lower limit) are read out in this order (total of NOP
×
4 values).
{numeric n
×
4+1}: Sweep parameter value
{numeric n
×
4+2}: Limit test result
0: FAIL
1: PASS
-1: Test was off.
{numeric n
×
4+3}: Upper border value of the limit
{numeric n
×
4+4}: Lower border value of the limit
Where, n is an integer between 0 and NOP-1.
Corresponding
key
No front panel key is available to execute this function.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...