480
Appendix B
Status Reporting System
Status Register Structure
Executing “*CLS” command on page 258 will clear all bits of the instrument status
register.
Executing “*CLS” command will clear all bits of the operation status condition register.
Table B-3
Status Bit Definitions of the Instrument Status Register
Bit
Name
Description
0
Single or Number of Groups
Sweep Complete
“1” is set when a single or group sweep was completed
after the last read of the register.
1
Service Routine Waiting or
Manual Trigger Waiting
1.
“1” is set when execution of an internal service
routine is completed or when the routine is put into
waiting state for operator’s response.
2.
“1” is set when the 4294A is put into the manual
trigger mode or GPIB/LAN trigger mode and is
waiting for a manual trigger.
2
Data Entry Complete
“1” is set when the terminator key is pressed.
3
Limit Failed, Trace B
“1” is set when a limit test failed on trace B.
4
Limit Failed, Trace A
“1” is set when a limit test failed on trace A.
5
Search Failed, Trace B
“1” is set when no value was obtained for a target of
marker search on trace B.
6
Search Failed, Trace A
“1” is set when no value was obtained for a target of
marker search on trace A.
7
Point Measurement Complete
*1
“1” is set when a measurement of a single point is
completed in sweep measurement. (when using the point
trigger)
8
Calibration or Compensation
Complete
“1” is set when each measurement data is completely
obtained for the adapter setting, user calibration, and
fixture compensation.
*1.This bit is set only when both of bits associated with the service request enable
register and the instrument event status enable register are enabled.
Table B-4
Status Bit Definitions of the Operation Status Condition Register
Bit
Name
Description
9
Printing
“1” is set while data is being transferred to the printer.
14
Program Running
“1” is set while an HP Instrument BASIC program is
running.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...