36
Chapter 3
Setting Measurement Conditions
Setting Measurement Parameters
Setting Measurement Parameters
To set measurement parameters, use the command given below. This command sets
measurement parameters for both trace A and trace B at the same time.
•
“MEAS” on page 348
You can set measurement parameters in the combinations shown in Table 3-1 depending on
the parameters used at the time of sending a command.
Table 3-1
Measurement parameters
Parameters for
sending a
command
Parameters to be set
Trace A
Trace B
IMPH
|Z|
θ
IRIM
R
X
LSR
Ls
Rs
LSQ
Ls
Q
CSR
Cs
Rs
CSQ
Cs
Q
CSD
Cs
D
AMPH
|Y|
θ
ARIM
G
B
LPG
Lp
G
LPQ
Lp
Q
CPG
Cp
G
CPQ
Cp
Q
CPD
Cp
D
COMP
Z
Y
IMLS
|Z|
Ls
IMCS
|Z|
Cs
IMLP
|Z|
Lp
IMCP
|Z|
Cp
IMRS
|Z|
Rs
IMQ
|Z|
Q
IMD
|Z|
D
LPR
Lp
Rp
CPR
Cp
Rp
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...