TOPV
TOPV
Syntax
TOPV <numeric>[OHM|DEG|RAD|SIE|H|F|PCT]
TOPV?
Description
Sets the maximum value of the display screen in the Y axis (vertical axis) direction (value
of the top of the grid).
If the difference, between this maximum value and the minimum value in the Y-axis
direction on the display screen (set with the “BOTV” command on page 273), is less than
10E-15 (for the log Y axis format, the ratio between the maximum value and the minimum
value is less than 5), the minimum value is automatically changed so that it is less than the
maximum value minus 10E-15 (for the log Y axis format, the value at which the ratio
between the maximum value and the minimum value is 5). Also, the single grid tick width
(SCALE/DIV) and the reference line value (REFERENCE VALUE) are automatically
changed so that they match the setting of the minimum value/maximum value in the Y-axis
direction.
In the log Y axis format, if the sign differs from that of the display screen minimum value,
the sign of the minimum value is automatically changed to the same sign as the maximum
value.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
key
[Scale Ref]
-
TOP VALUE
<numeric>
Description
Maximum value of the display screen in the Y axis (vertical axis)
direction
Range
Varies depending on the measurement parameter as follows:
For |Z|, R, X: -1E12 to 1E12
For other than |Z|, R, X: -1E9 to 1E9
Initial value
Varies depending on the measurement parameter as follows:
For |Z|, R, X, Rs, Rp: 1E6
For
θ
: 180
For |Y|, G, B, D: 1
For Cs, Cp: 1E-3
For Ls, Lp: 10
For Q: 1E3
Unit
Varies depending on the measurement parameter (refer to the
explanation of unit in “SCAL” on page 423).
Resolution
1E-15
*1
*1.This is the minimum value (when the set value is small). The resolu-
tion becomes larger as the set value becomes larger.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...