392
Chapter 16
GPIB Command Reference
OUTPRESR?
OUTPRESR?
Syntax
OUTPRESR?
Description
If the sweep parameter is frequency and the measurement parameter is |Z|-
θ
, within the
waveform analysis range specified with the “ANARANG” command on page 267,
analyzes the resonant point and ripple, and reads out the result. If the sweep parameter is
not frequency or if the measurement parameter is not |Z|-
θ
, executing this command causes
an error and all readouts are 0. (Query only)
Query response
{numeric 1},{numeric 2},{numeric 3},{numeric 4},{numeric 5},{numeric 6},{numeric 7}<newline><^END>
The readouts are as follows:
{numeric 1}:
Value of Z
r
(resonant impedance).
{numeric 2}:
Value of f
r
(resonant frequency).
{numeric 3}:
Value of Z
a
(anti-resonant impedance).
{numeric 4}:
Value of f
a
(anti-resonant frequency).
{numeric 5}:
Rpl
1
(the maximum value of the left-side ripple (the sweep parameter
difference between the peak and the left adjacent negative peak)
within the range to the left of the resonant point within the waveform
analysis range).
{numeric 6}:
Value of Rpl
2
(the maximum value of the right ripple (the sweep
parameter difference between the peak and the right adjacent negative
peak) within the range between the resonant point and the
anti-resonant point within the waveform analysis range).
{numeric 7}:
Rpl
3
(the maximum value of the left-side ripple within the range to the
right of the resonant point within the waveform analysis range).
This command searches for phase 0° points within the analysis range from the left edge,
and defines the 1st detected point as the resonant point and the 2nd detected point as the
anti-resonant point. Even if 3 or more phase 0° points are detected, only the first 2 points
are used. If only 1 phase 0° point is detected, 0 is read out for parameters other than Z
r
, f
r
,
and Rpl
1
. If no phase 0° point is detected, 0 is read out for all the parameters.
The peak and negative peak used at the execution of this command satisfies the condition
set with the “THRR” command on page 452. They are not related to the peak definition in
the marker search function.
Corresponding
key
No front panel key is available to execute this function.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...