
Chapter 4
63
Preparing For Accurate Measurement
Fixture Compensation
1. Ch
apter
T
itle
2
. Cha
p
te
r T
itle
3
. Chap
te
r T
itle
4
. Gettin
g Read
y Fo
r
5. Chap
te
r T
itle
Accu
ra
te
M
e
as
ur
em
e
n
t
Fixture Compensation
To turn on or off the open/short/load compensation of the fixture compensation functions,
use the command given below. You can set open/short/load compensations independently.
However, if no compensation coefficients are retained, the functions cannot be turned on.
•
“COMST{A|B|C}” on page 282
To measure open data, short data and load data for calculating compensation coefficients in
open/short/load compensation of the fixture compensation functions, use the commands
given below. Executing these commands calculates fixture compensation coefficients of
measured data and stores them in volatile memory (RAM). In addition, the compensation
function for measured data is turned on.
•
“COM{A|B|C}” on page 281
NOTE
User calibration data and fixture compensation data are measured either at fixed points
preset by the Agilent 4294A (fixed measurement points) or at sweep measurement points
set at the time of command execution (user-selected measurement points). In the case of
the former, the compensation coefficient at a sweep measurement point is obtained by
interpolation. To select points for measuring data, use the “CALP” command on page 275.
To set standard values to be used in measuring these data, use the commands given below.
•
“DCOMOPEN{G|C}” on page 293
•
“DCOMSHOR{R|L}” on page 294
•
“DCOMLOAD{R|L}” on page 292
To read open data, short data or load data measured, use the command given below.
•
“OUTPCOMC{1-3}?” on page 374
You can calculate fixture compensation coefficients by entering open data, short data and
load data (data read by the above or similar command). To enter data, use the command
given below. Executing these commands calculates the fixture compensation coefficients
of measured data and stores them in volatile memory (RAM). In addition, the
compensation function for measured data is turned on.
•
“INPUCOMC{1-3}” on page 321
NOTE
Enter data after setting the standard values used in measuring the data because these values
are needed to calculate fixture compensation coefficients.
NOTE
Enter data after making settings identical to the data measurement points (fixed
measurement points, user-selected measurement points) set at the time when the input data
were acquired. In addition, in the case of the user-selected measurement points, enter data
after making the sweep settings identical to the data measurement points.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...