Chapter 4
Preparing For Accurate Measurement
Adapter Setting
Adapter Setting
To select an adapter, use the command given below.
•
“E4TP” on page 309
If you select any adapter other than NONE, you need to measure the data indicated in the
table below to calculate the setup data.
To measure data, use the command given below.
•
“ECAL{P|A|B|C}” on page 309
After measuring the necessary data, execute the command given below to calculate the
setup data for the selected adapter and then store this in the nonvolatile memory
(EEPROM).
•
“ECALDON” on page 309
NOTE
If you execute the “ECAL{P|A|B|C}” command to measure unnecessary data, an error
occurs and the command is ignored.
To suspend the above process, execute the command given below.
•
“ECALQUI” on page 310
A sample program for adapter setting is shown in Example 4-1. This program is given the
file name adapter.bas and is stored on the sample program disk.
This program selects 7mm 42942A for adapter setting, measures phase data, open data,
short data and load data and then stores them in nonvolatile memory.
Lines 40 to 100
These lines distinguish between the external controller and Instrument
BASIC and set the GPIB address and the select code.
Line 110
This line substitutes the adapter selection to the variable Adapter$.
Lines 130 to 160 These lines reset the Agilent 4294A and then set the adapter selection
to the variable Adapter$.
Lines 180 to 190 These lines set 1 to Bit 8 of the instrument event status register (the bit
for completion of the measurement of calibration data and
compensation data) and to Bit 2 of the service request enable register
so that SRQ can be used.
Adapter
Phase
Open
Short
Load
NONE
Unnecessary
Unnecessary
Unnecessary
Unnecessary
4TP 1M
Necessary
Unnecessary
Unnecessary
Necessary
4TP 2M
Necessary
Unnecessary
Unnecessary
Necessary
7mm 42942A
Necessary
Necessary
Necessary
Necessary
PROBE 42941A
Necessary
Necessary
Necessary
Necessary
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...