438
Chapter 16
GPIB Command Reference
SMKR{1-7}
SMKR{1-7}
Syntax
SMKR{1|2|3|4|5|6|7} {ON|OFF|1|0}
SMKR{1|2|3|4|5|6|7}?
Description
For the active trace (set with the “TRAC” command on page 457), turns on/off a sub
marker (marker number 1 to 7). In the coupled marker mode (set with the “MKRCOUP”
command on page 352), the setting is applied to both the A and B traces regardless of the
active trace setting. The sub marker set to on with this command is displayed at the same
position as the marker. If the marker function is off, executing this command causes an
error and the command is ignored.
Parameters
Query response
{ON|OFF|1|0}<newline><^END>
Corresponding
key
[Marker]
-
SUB MKR
- {
1: OFF
|
2: OFF
|
3: OFF
|
4: OFF
|
5: OFF
|
6: OFF
|
7: OFF
}
[Marker]
-
CREAR
SUB MKR
- {
1: OFF
|
2: OFF
|
3: OFF
|
4: OFF
|
5: OFF
|
6: OFF
|
7: OFF
}
SMKRAUV{1-7}?
Syntax
SMKRAUV{1|2|3|4|5|6|7}?
Description
For the active trace (set with the “TRAC” command on page 457), if the measurement
parameter is a vector value (for COMPLEX Z-Y), reads out the subsidiary readout of the
measurement parameter value at the sub marker position (marker number 1 to 7). (Query
only)
Query response
{numeric}<newline><^END>
If the measurement parameter is a scalar value (for other than COMPLEX Z-Y), 0 is
always read out. If the measurement parameter is a vector value (for COMPLEX Z-Y), the
readout is as follows:
•
If the polar chart format is selected, the phase is read out.
•
If the complex plane format is selected, reactance (for trace A) or susceptance (for trace
B) is read out.
Corresponding
key
No front panel key is available to execute this function.
Description
ON or 1
Turns on a sub marker.
OFF or 0 (initial value)
Turns off a sub marker.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...