460
Chapter 16
GPIB Command Reference
WIDFVAL
WIDFVAL
Syntax
WIDFVAL <numeric>[OHM|DEG|RAD|SIE|H|F|PCT]
WIDFVAL?
Description
For the active trace (set with the “TRAC” command on page 457), if the free cutoff value
setting is enabled (specified to FIXed with the “WIDVTYPE” command on page 462), sets
a cutoff point. If
Δ
maker is on, the cutoff value is set the result of subtracting the specified
value from the measurement parameter value of
Δ
maker position.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
key
[Search]
-
WIDTHS DEF MENU
-
FIXED VALUE
WIDSIN
Syntax
WIDSIN
Description
For the trace bandwidth analysis function for the active trace (set with the “TRAC”
command on page 457), further searches for another cutoff point inside the already
detected cutoff point. If no cutoff point is detected, the message indicating the result is
displayed (no error occurs). If the trace bandwidth analysis function is off, executing this
command causes an error and the command is ignored. (No query)
Corresponding
key
[Search]
-
WIDTHS DEF MENU
-
SEARCH IN
<numeric>
Description
Cutoff value
*1
Range
-100E6 to 100E6
Initial value
0
Unit
Varies depending on the measurement parameter (refer to the
explanation of unit in “SCAL” on page 423).
*1.If the measurement parameter is a vector value (for COMPLEX Z-Y)
and if the polar chart format is selected, set the value in amplitude. If
the complex plane format is selected, set the value in resistance (for
trace A) or conductance(for trace B).
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...