196
13
Application Sample Programs
Measuring Dielectric Material
air capacitor is used as the LOAD standard.
Lines 2230 to 2260: Waits until you adjust the distance between
the electrodes so that the limit test results display PASS (an
appropriate air capacitor value is obtained) and press the
Continue
key.
Lines 2270 to 2300: Checks the limit test result.
Lines 2330 to 2340: Retrieves the measurement result (Cp) and
stores it into the Load_c variable for the LOAD standard value.
Lines 2360 to 2410: Sets up the measurement conditions used
when performing fixture compensation.
Lines 2420 to 2470: Defines the values of each
OPEN/SHORT/LOAD standard.
NOTE
In order to define the LOAD standard using the values Cp and G, the OPEN standard and
LOAD standard must be defined inversely.
Lines 2490 to 2640: Measures the compensation data for each of
OPEN/SHORT/LOAD.
NOTE
Because the OPEN and LOAD standards are defined inversely, the commands for
measuring the OPEN and LOAD compensation data are also used inversely.
Lines 2690 to 2730 The Input_t subroutine.
Lines 2750 to 2890 The Input_d subroutine.
Lines 2270 to 2820: When you use electrode A or B, sets the
electrode diameter from the values already set in the variables
within the program.
Lines 2830 to 2870: When you use electrode C or D, sets the
electrode diameter from user input.
Lines 2910 to 3100 The Sweep subroutine.
Lines 2920 to 2930: Waits until you insert the MUT into the
16451B and press the
Continue
key.
Lines 2940 to 3000: Measures Cp and D.
Lines 3010 to 3090: Retrieves the measurement results and stores
them into the Cp_data and D_data arrays.
Lines 3120 to 3230 The Calc subroutine. Using the Cp_data and D_data arrays, calculates
the complex relative permittivity and the dielectric dissipation factor.
Lines 3250 to 3350 Sets the softkey labels and the subroutines to be executed when each
key is pressed. Then, waits until you press one of the softkeys.
Lines 3370 to 3580 Processing when the softkey labeled as “er' - tan d” is pressed.
Displays the real part of the complex relative permittivity (
) as
trace A and the dielectric dissipation factor as trace B.
Lines 3600 to 3810 Processing when the softkey labeled as “|er| - tan d” is pressed.
Displays the magnitude of the complex relative permittivity
ε
r
'
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...