Chapter 16
321
GPIB Command Reference
INPUCOMC{1-3}
16. GPI
B
Com
m
an
d Re
fere
nce
( F -
N )
INPUCOMC{1-3}
Syntax
INPUCOMC{1-3} <numeric 1>,<numeric 2>,..,<numeric NOP
×
2-1>,<numeric NOP
×
2>
Description
Enters the OPEN/SHORT/LOAD compensation data for fixture compensation, calculates
the compensation coefficient, and stores it into the non-volatile memory (backup memory).
The “INPUCOMC1” command is for OPEN compensation data (G-B); the
“INPUCOMC2” command, SHORT compensation data (R-X); the “INPUCOMC3”
command, for LOAD compensation data (R-X).
Enter data after setting the standard values, because the standard values to be used in
measuring data is needed to calculate fixture compensation coefficients.
A mismatch, between the number of required data at the execution of the command
(defined by the measurement point setting for fixture compensation data (set with the
“CALP” command on page 275) and the number-of-points setting) and the number of
entered parameters, causes an error and the command is ignored. (No query)
Parameters
Total of NOP
×
2 parameters are required because the fixture compensation data value
(complex number) of each measurement point is divided into 2 parameters for its real part
and imaginary part. Where, NOP is the number of points (for user measurement points) or
100 (for fixed measurement points), and n is an integer between 1 and NOP.
Corresponding
key
No front panel key is available to execute this function.
INPUDATA
Syntax
INPUDATA <numeric 1>,<numeric 2>,..,<numeric NOP
×
2-1>,<numeric NOP
×
2>
Description
Enters a data array (refer to “Internal data arrays” on page 81). A mismatch, between the
number-of-points setting at the execution of the command and the number of entered
parameters, causes an error and the command is ignored. (No query)
Parameters
Total of NOP
×
2 parameters are required because the measurement parameter value of each
measurement point (complex number) is divided into 2 parameters for its real part and
imaginary part. Where, NOP is the number of points, and n is an integer between 1 and
NOP.
Corresponding
key
No front panel key is available to execute this function.
Description
<numeric n
×
2-1>
Real part of the compensation data value (complex number) of the
n-th measurement point.
<numeric n
×
2>
Imaginary part of the compensation data value (complex number) of
the n-th measurement point.
Description
<numeric n
×
2-1>
Real part of the measurement parameter value (complex number) of
the n-th measurement point.
<numeric n
×
2>
Imaginary part of the measurement parameter value (complex
number) of the n-th measurement point.
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...