Application Sample Programs
Measurement using scanner
1. Ch
apter
T
itle
2.
Ch
ap
te
r
T
itl
e
1
3
. Application Sample
4
. Chap
te
r T
itle
5. Chap
te
r T
itle
Programs
performed here. Because the phase compensation is common to all the channels, perform it
for one channel. For the phase compensation, you need to bring the measurement terminal
into the open state as shown in Figure 13-10.
Figure 13-10
Open state
Step 3.
Setting the measurement conditions
Set the list sweep table. Set the sweep condition for each channel for each segment. In this
sample program, the sweep conditions from channels 0 through 3 are assigned to segments
1 through 4.
Step 4.
Reading out the compensation coefficient of the fixture compensation
For each channel, perform the open, short and load fixture compensations. For the load
compensation, use a working standard whose value is known. Data measurement for the
compensation coefficient is performed for all segments. After calculating the
compensation coefficient, read out the compensation coefficient for each channel.
Step 5.
Creating the scan-aware compensation coefficient
From the compensation coefficient array read out in Step 4, combine the segments that
correspond to individual channels to create the scan-aware compensation coefficient.
Step 6.
Executing impedance measurement
Change the sweep range of the manual sweep and the channels and perform impedance
measurement for the multiple channels.
Sample program
Example 13-6 shows a sample measurement program assuming that it is used in a scanning
system (4 channels). This program is stored in the sample program disk as scan.bas.
Lines 70 to 170
Identifies the external controller and Instrument BASIC and sets the
GPIB address and select code.
Lines 200 to 230 Prepares data for the list sweep table using DATA statements.
Segments correspond to channels 0 to 3 from the top.
Lines 250 to 300 Assigns the measurement conditions (adapter setting: 2-m test lead,
measurement parameter: |Z|-q, sweep type: LIST, oscillator power
level setting unit: VOLT, and DC bias setting unit: VOLT) into the
Summary of Contents for 4294A
Page 4: ...4 ...
Page 18: ...18 Contents ...
Page 24: ...24 Chapter1 To make effective use of this manual Other manuals attached to this unit ...
Page 94: ...94 Chapter6 Reading Writing Measurement Data Reading Writing Data ...
Page 110: ...110 Chapter7 Processing Measurement Results Analyzing Measurement Results ...
Page 136: ...136 Chapter10 Handling Errors Sample program for error handling ...
Page 246: ...246 13 Application Sample Programs File Transfer Function ...
Page 254: ...254 Chapter15 Setting the Display LCD Sample program for setting the LCD screen ...
Page 507: ...Index Index 507 Index ...
Page 508: ...508 Index Index ...
Page 509: ...Index Index 509 Index ...
Page 510: ...510 Index Index ...
Page 511: ...Index Index 511 Index ...
Page 512: ...512 Index Index ...
Page 513: ...Index Index 513 Index ...
Page 514: ...514 Index Index ...
Page 515: ...Index Index 515 Index ...
Page 516: ...516 Index Index ...
Page 517: ...Index Index 517 Index ...
Page 518: ...518 Index Index ...