PRF Mode
You can choose between the
Auto
and the
Manual
mode for the PRF (pulse repetition frequency) setting.
When you select
Auto
the function
is
locked. The PRF is set automatically to a value of 30%
of the maximum possible PRF.
Select
Manual
to set the
PRF Value
The pulse repetition frequency indicates the number of
times an initial pulse is triggered per second.
The larger your test object, the smaller PRF values are
needed to avoid phantom echoes. However, the A-scan
update rate becomes lower in the case of smaller PRF
values. High values are therefore required if a test ob-
ject should be scanned fast.
The best way to determine the suitable PRF value is by
experimenting: Start from the highest value and reduce
the value until there are no more phantom echoes.
Pulser Mode
The USM 100 can be equipped with an optional spike
pulser. If the function
is available, you can
choose between the
Spike
and the
Square
pulser type.
When the spike pulser is selected here, the functions
and
are locked. These parame-
ters are then set automatically and optimized for high
frequency probes.
Phantom Echo
With the function
you can switch on the
phantom echo detection. When the function is switched
on, detected phantom echoes become visible by caus-
ing the A-scan to jump. If no phantom echoes are de-
tected, no difference is visible in the A-scan display.
Phantom echoes can occur depending on the material
properties and the set pulse repetition frequency, espe-
cially if the pulse repetition frequency is too high.
Содержание Krautkramer USM 100
Страница 1: ...Krautkr mer USM 100 Operating Manual Id No 160M4395 Rev 9 09 2023...
Страница 14: ...14 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 21: ...Krautkr mer USM 100 Rev 9 09 2023 21 Introduction 1...
Страница 36: ...36 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 37: ...Krautkr mer USM 100 Rev 9 09 2023 37 Standard package and accessories 2...
Страница 45: ...Krautkr mer USM 100 Rev 9 09 2023 45 Initial start up 3...
Страница 66: ...66 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 67: ...Krautkr mer USM 100 Rev 9 09 2023 67 Principles of operation 4...
Страница 103: ...Krautkr mer USM 100 Rev 9 09 2023 103 Operation 5...
Страница 238: ...238 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 239: ...Krautkr mer USM 100 Rev 9 09 2023 239 Documentation 6...
Страница 257: ...Krautkr mer USM 100 Rev 9 09 2023 257 Maintenance and care 7...
Страница 267: ...Krautkr mer USM 100 Rev 9 09 2023 267 Interfaces and Peripherals 8...
Страница 287: ...Krautkr mer USM 100 Rev 9 09 2023 287 Appendix 9...
Страница 303: ...Krautkr mer USM 100 Rev 9 09 2023 303 Specifications 10...
Страница 309: ...Krautkr mer USM 100 Rev 9 09 2023 309 Index 11...
Страница 329: ...Krautkr mer USM 100 Rev 9 09 2023 329 empty page...