Remote control
8 Interfaces and Peripherals
Krautkrämer USM 100
275
AG
flag
dec
Perform the
AutoXX
gain adjustment on the specified gate.
The system gain is adjusted so that the peak signal in the
specified gate is set to 80 % screen height or a user
specified amplitude.
AG <s t>
where
s = A or a for gate A
B or b for gate B
t = 10 … 100 (1 count = 1 % screen height,
optional user specified amplitude)
The message
ERROR: Automatic Gain Adjustment
Failed
indicates that the dynamic range of the system gain
can't achieve the target amplitude.
AM
dec
Read or set the gate A logic.
AM <n>
where
n = 0 for Off, 1 for Positive, 2 for Negative.
Code
Data type
Description
Function
Page
Содержание Krautkramer USM 100
Страница 1: ...Krautkr mer USM 100 Operating Manual Id No 160M4395 Rev 9 09 2023...
Страница 14: ...14 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 21: ...Krautkr mer USM 100 Rev 9 09 2023 21 Introduction 1...
Страница 36: ...36 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 37: ...Krautkr mer USM 100 Rev 9 09 2023 37 Standard package and accessories 2...
Страница 45: ...Krautkr mer USM 100 Rev 9 09 2023 45 Initial start up 3...
Страница 66: ...66 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 67: ...Krautkr mer USM 100 Rev 9 09 2023 67 Principles of operation 4...
Страница 103: ...Krautkr mer USM 100 Rev 9 09 2023 103 Operation 5...
Страница 238: ...238 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 239: ...Krautkr mer USM 100 Rev 9 09 2023 239 Documentation 6...
Страница 257: ...Krautkr mer USM 100 Rev 9 09 2023 257 Maintenance and care 7...
Страница 267: ...Krautkr mer USM 100 Rev 9 09 2023 267 Interfaces and Peripherals 8...
Страница 287: ...Krautkr mer USM 100 Rev 9 09 2023 287 Appendix 9...
Страница 303: ...Krautkr mer USM 100 Rev 9 09 2023 303 Specifications 10...
Страница 309: ...Krautkr mer USM 100 Rev 9 09 2023 309 Index 11...
Страница 329: ...Krautkr mer USM 100 Rev 9 09 2023 329 empty page...