5 Operation
Calibration
166
Krautkrämer USM 100
Multi BW
With
Multi BW
(multi backwall), the calibration only
needs one reference value
S Ref 1
. The first and the
second backwall echo are generated on a single cali-
bration block.
During the calibration procedure the
adjusted automatically.
– Select the
2-Point Cal.
panel.
– Tap in the A-scan and tap the
to display the function group (see page 155).
– Select
and select
Multi BW
.
– Select
S Ref 1
and set the value according to the cal-
ibration block used.
– Couple the probe to the calibration block. The gates
are set automatically according to the value of
S Ref 1
.
– In the Command bar tap on
(see page 74)
to execute the calibration procedure.
Содержание Krautkramer USM 100
Страница 1: ...Krautkr mer USM 100 Operating Manual Id No 160M4395 Rev 9 09 2023...
Страница 14: ...14 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 21: ...Krautkr mer USM 100 Rev 9 09 2023 21 Introduction 1...
Страница 36: ...36 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 37: ...Krautkr mer USM 100 Rev 9 09 2023 37 Standard package and accessories 2...
Страница 45: ...Krautkr mer USM 100 Rev 9 09 2023 45 Initial start up 3...
Страница 66: ...66 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 67: ...Krautkr mer USM 100 Rev 9 09 2023 67 Principles of operation 4...
Страница 103: ...Krautkr mer USM 100 Rev 9 09 2023 103 Operation 5...
Страница 238: ...238 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 239: ...Krautkr mer USM 100 Rev 9 09 2023 239 Documentation 6...
Страница 257: ...Krautkr mer USM 100 Rev 9 09 2023 257 Maintenance and care 7...
Страница 267: ...Krautkr mer USM 100 Rev 9 09 2023 267 Interfaces and Peripherals 8...
Страница 287: ...Krautkr mer USM 100 Rev 9 09 2023 287 Appendix 9...
Страница 303: ...Krautkr mer USM 100 Rev 9 09 2023 303 Specifications 10...
Страница 309: ...Krautkr mer USM 100 Rev 9 09 2023 309 Index 11...
Страница 329: ...Krautkr mer USM 100 Rev 9 09 2023 329 empty page...