Pulser Receiver
5 Operation
Krautkrämer USM 100
135
5.11 Pulser Receiver
In this function group you will find all functions for setting
the pulser and the receiver.
Voltage
If the square wave pulser is chosen as pulser mode (see
page 139), then you can set the pulser voltage within
the range of 100 … 350 V in steps of 10 V.
ATTENTION
Use the data sheet for your probe to check
which maximum voltage is allowed to be ap-
plied.
Note
The
and the
can be
automatically limited, depending on the
pulse repetition frequency mode (see
page 139). This function helps to avoid heat
accumulations in the pulser electronics.
Содержание Krautkramer USM 100
Страница 1: ...Krautkr mer USM 100 Operating Manual Id No 160M4395 Rev 9 09 2023...
Страница 14: ...14 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 21: ...Krautkr mer USM 100 Rev 9 09 2023 21 Introduction 1...
Страница 36: ...36 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 37: ...Krautkr mer USM 100 Rev 9 09 2023 37 Standard package and accessories 2...
Страница 45: ...Krautkr mer USM 100 Rev 9 09 2023 45 Initial start up 3...
Страница 66: ...66 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 67: ...Krautkr mer USM 100 Rev 9 09 2023 67 Principles of operation 4...
Страница 103: ...Krautkr mer USM 100 Rev 9 09 2023 103 Operation 5...
Страница 238: ...238 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Страница 239: ...Krautkr mer USM 100 Rev 9 09 2023 239 Documentation 6...
Страница 257: ...Krautkr mer USM 100 Rev 9 09 2023 257 Maintenance and care 7...
Страница 267: ...Krautkr mer USM 100 Rev 9 09 2023 267 Interfaces and Peripherals 8...
Страница 287: ...Krautkr mer USM 100 Rev 9 09 2023 287 Appendix 9...
Страница 303: ...Krautkr mer USM 100 Rev 9 09 2023 303 Specifications 10...
Страница 309: ...Krautkr mer USM 100 Rev 9 09 2023 309 Index 11...
Страница 329: ...Krautkr mer USM 100 Rev 9 09 2023 329 empty page...